PEM (Center)
takaaki.yokozawa - 13:55 Sunday 06 April 2025 (33273) Print this report
PEM injection 250406
I performed the PEM injection around PR3
related klog : klog33254

2025/04/06 13:00:00 - 13:31:00 Shaker injection PR3 #1
Excitation : K1:PEM-EXCITATION_MCF0_RACK_7_EXC
50 - 200 Hz , 1Hz resolution, 10 s in each measurement, 100 cnt

2025/04/06 13:00:00 - 13:31:00 Shaker injection PR3 #2
Excitation : K1:PEM-EXCITATION_MCF0_RACK_8_EXC
50 - 200 Hz , 1Hz resolution, 10 s in each measurement, 100 cnt
During this measurement, locked loss happened.
Comments to this report:
tatsuki.washimi - 12:28 Tuesday 08 April 2025 (33304) Print this report

I analyzed the data of the shaker injection at PR3 ground.

At the time of the around 192Hz injection, the strain spectrum was noisy.

The projection, by assuming these noises were caused by the injection, was much smaller than the current sensitivity.

Images attached to this comment
tatsuki.washimi - 13:30 Tuesday 08 April 2025 (33305) Print this report

I also analyzed the data of the shaker injection at the PR3 beam-duct.

Similar to the ground shaking, the strain spectrum around 80Hz got noisy when the shaker vibrated around 190Hz (and around 173Hz).

At 89 Hz, a liner-like response was found.

 

Images attached to this comment
takaaki.yokozawa - 8:53 Wednesday 09 April 2025 (33316) Print this report
I tried to inject the 192 Hz to same shaker 8:47 - 8:52 today, but no significant excess can be seen in DARM signal.
So, that excess would be the chance coincidence of glich which appeared in other reason.
Images attached to this comment