takaaki.yokozawa - 8:55 Friday 04 April 2025 (33236) Print this report
Measurement of the resonant peaks 45 Hz pitch in Type-A suspension
I tried to identify which suspension had the resonant peak of 45 Hz in pitch for the Type-A suspensions.
I injected 40 - 50 Hz white noise to MNP test filter bank. This is the same method with klog32983

Fig.1. IX about 45 Hz
Fig.2. IY about 44 Hz
Fig.3. EX about 44.5 Hz
Fig.4. EY about 44.4 Hz (cannot be seen in silent case)
Images attached to this report
Comments to this report:
takaaki.yokozawa - 8:59 Friday 04 April 2025 (33239) Print this report
I generated the NB filter for the IX (45 Hz) using P04
I will check the performance of this filter after locking the interferometer
Images attached to this comment
takaaki.yokozawa - 12:07 Saturday 05 April 2025 (33266) Print this report
I implemented the 45 Hz pitch damping filter for IY and EX.
(We cannot see the peak in EY, so I didn't perform it)
I implemented those filters in P4 NBDAMP

Fig.1. and Fig.2. showed the transfer function for IY and EX.

Fig.3. showed the spectrum of DARM in the state of the INCREASING_LAS_POWER (red)with and (blue)without damping filters.
Actually, the damping of peaks can be succeeded by checking the spectrum of DARM and DHARD_P, but the spectrum around the peaks significantly increased.
So, we need to perform more narrow band filter and/or reducing the gain.
Images attached to this comment
takaaki.yokozawa - 14:16 Sunday 06 April 2025 (33274) Print this report
I found those filters cannot be used in OMC DC readout due to unknown reason.
Started the oscillation and OMC DC PDs saturated after ~10 min.
yuta.michimura - 17:55 Sunday 06 April 2025 (33275) Print this report

It seems like DHARD_Y has better SNR than DHARD_P to see peaks around 45 Hz. Interestingly, POS_SPOL also sees this motion (could be due to ITM birefringence). See attached. These channels would be useful for noise subtraction as well (sadly, they are 256 Hz DQ channels). See attached for noise subtration result using DHARD_Y with simple bandpass filters around these peaks.

Images attached to this comment
Non-image files attached to this comment
takaaki.yokozawa - 8:41 Tuesday 08 April 2025 (33296) Print this report
I turned the NBfilter gain and band pass frequency to perform the damping with the OMC DC readout.
Fig.1. - Fig.3. showed the filter performance of each suspension (IX, IY and EX)

Actually, when we performed the NB damp (P4) to each suspension, the peak itself can be damped, but still increased the sideband and no improvement of the sensitivity can be found(Fig.4. and Fig.5.).
Images attached to this comment