Reports 1-1 of 1 Clear search Modify search
PEM (Center)
tatsuki.washimi - 14:59 Thursday 03 April 2025 (33216) Print this report
Comment to PEM injection test 250403 (33202)

I analyzed the OMC acoustic injection data:

2025/04/03 05:31:00 - 05:51:00 Speaker injection OMC area
Excitation : K1:PEM-EXCITATION_SR3_RACK_6_EXC
50 - 900 Hz white noise injection 1,500 cnt

  • Fig.1: Spectrograms for the strain, the microphone, the geophone, the OMC error signal, and the square of the OMC error signal
  • Fig.2: ASDs of these channels, for the reference and the injected time
  • Fig.3: Microphone -> Geophone coupling and projection
  • Fig.4: Microphone -> OMC error coupling and projection
  • Fig.5: Geophone -> OMC error coupling and projection

 

Because any significant excess was found in the OMC error signal limited by its sensing noise, a noise projection on the strain was impossible.

Images attached to this comment
Search Help
×