Reports 1-1 of 1 Clear search Modify search
CAL (Gcal general)
Kohei Mitsuhashi - 23:41 Tuesday 01 April 2025 (33187) Print this report
3D measurement of Ncal weights

Abstract

 
We requested 3D measurement of Ncal weights, and the measurment was done.
 

Details


We requested the Mecdidhanical Engineering Center(機械工学センター) at KEK to conduct 3D measurement using Coordinate Measuring Machine (CMM) of twelve copper weights (Ø50.1mm, height 50mm)in picture 2 used for NCal and one aluminum weight (Ø50.1mm, height 50mm) in picture 3 made from excess material left over during the fabrication of the NCal disk. The measurements were performed using a UPMC 850 (ZEISS) in picture 1.
The measurement was conducted using a contact probe, employing the radial method. Since the aluminum weight was too light, a fixture was used to prevent it from moving when touched by the probe. Due to this setup, the measurable points on the aluminum weight were limited, making it impossible to calculate roundness. Instead, the standard deviation was determined.
And roundness were measured at two point. Point are 10 mm and 40 mm from the floor of UPMC 850.

The results are as follows.
To prevent from using wrong terms, I wrote down a tabel in japanese. Sorry.
 
 
 
ナンバリング直径平行度真円度_h-40真円度_h-10温度(°C)
150.07400.04680.01280.009422.2
250.05820.04460.03570.015722.5
350.08260.05560.01380.008122.5
450.09990.06580.01790.008722.4
550.09990.05390.01690.010322.7
650.07270.10720.01640.016722.5
750.08510.05000.01820.011822.4
850.14560.09580.01020.013422.5
950.08560.04230.00680.009522.5
1050.04860.05480.00840.013722.2
1150.13320.07140.01640.009322.3
1250.10860.05940.00690.010222.4
13(Ai)50.04980.05600.0328 *0.0378*22.2
* the standard deviation
 
 
Images attached to this report
Comments to this report:
dan.chen - 15:26 Wednesday 02 April 2025 (33191) Print this report

Here are some comments regarding this report:

  • In the original Ncal design, tungsten was intended to be used for the masses, but for this trial, copper was chosen due to its ease of machining and other practical reasons.
  • The copper masses measured this time are expected to be used in upcoming tests at KEK.
  • Whether the results of this measurement pose an issue in terms of uncertainty is currently under evaluation. (This is expected to be clarified as part of the ongoing calculation of Ncal’s uncertainty budget/requirements.)