Attached are trend graphs of the temperatures in the compressors of the cryocooler and Q-mass in the EYC.
The temperatures in the compressors temporarily rose due to the cooling water outage, but have now returned to normal values.
It was also confirmed that there was no change in the partial pressure of residual gas in the EYC after the cooling water was stopped.
Yea and Yer compressors stopped due to cooling water problems. Restarted after cooling water was restored.
P-53, P-54, and P-55 compressors continued operation, but alarms were displayed, so compressors were temporarily stopped and restarted in order to reset the displays.
As we were worried, the temp of EY 50K REFARM HEAD started increasing again because of the violation of the naive condition of cooling.
The temperature increases at EY 50K REFARM HEAD seemed to stop and start decreasing.
The outer shields' temperatures seemed to be below 157K (H2O outgassing temperature) even after one of the cryocoolers for the payload was off. On the other hand, the temperatures were over 157K if the cryocoolers for the payloads were not operated.