takaaki.yokozawa - 9:20 Friday 24 January 2025 (32443) Print this report
Finesse measurement Xarm 250124
I performed the FInesse measurement Xarm.
Before this measurement, I tried to check the possible power of IMC output.
We set the
IMC_LSC_LOCKED for IO guardian
IR_LOCKED for Xarm guardian
HWP of the REFL table set to 168 deg.
(HWP of the PSL was 13 deg.)
As shown in Fig.1. we can increase the IMC output power to 7.1 W in 30 deg.

We didn't care the alignment so much in this measurement.

ATE : 2025/01/23 22:47 UTC
ARM : X
TEMP ITM : 89.1
TEMP ETM : 89.3
NUM : 10
IFO REFL HWP : 168.0
PSL HWP : 30.0
IMC POWER : 7.2
VALUE : 1456.1
ERROR : 10.3

Measured data is stored to /users/Commissioning/data/Finesse/Xarm/20250123-224708

Summary of the Finesse (and temperature)
https://gwdet.icrr.u-tokyo.ac.jp/~controls/trendMon/finesse.html
Images attached to this report
Comments to this report:
shoichi.oshino - 14:37 Tuesday 13 January 2026 (36121) Print this report
I reanalysed past finesse measurement data with new fitting function that uses an offset value.
The results obtained by reanalysis were very close to the values ​​obtained after finesse was recovered(klog 35965).
Images attached to this comment