VAC (EY)
shinji.miyoki - 9:08 Thursday 16 January 2025 (32353) Print this report
GV EYA manual close

Yamamoto kun reported via slack,

  • @uchiyama@Nobuhiro KIMURA
     Now KGRVAC:CCG_Y30_01:VAL:PRESS reaches 1.2e-5Pa which is the current threshold of the interlock according to klog#32347. But GV at EYA is still open.

Fig.3 shows the X-30 pressure. I and Uchiyama-kun also confirmed that the GV_eya was open. Anyway Uchiyama-kun closed EYA GV by using Switchbot manually, although Uchiyama-kun wondered whether it should be closed or not because this enhancement should be the increase of nitrogen that is harmless for a mirror.

I checked the vacuum level and cryocooler status for EY as Fig.1 and Fig.2. Actually, the vacuum level at EYA CC-10(EYC side) was enhanced, but it started decreasing as in Fig.2. The temperatures also started decreasing as in Fig.1.

I would like to propose,

  • set trigger for GV closing: ~ 6*10^-5 Pa (near 10^-4 pa),
  • set the slack alert for enhancement of the vacuum level: 10^-5 Pa,

because,

  • The serious leak will result in immediate vacuum level enhancement over 10^-4 Pa according to our past experiences. In such a case, the alert time and the closing time are almost the same.
  • Slight leaks or Nitrogen releases from cryostats are not so serious for a short time (~ several hours), so we can have time to understand the situation and judge what we should do.
Images attached to this report
Comments to this report:
shinji.miyoki - 10:25 Thursday 16 January 2025 (32354) Print this report

Nakagaki-san reported that,

  • a GV will be closed if the pressure level "OVERs", not "Reaches" the setting value!

So, no closing at 12.x10^-5 Pa that is the setting value was no problem.

However, I asked Kimura-san to set new trigger values for GV closing at 5.9*10^-5 Pa.

nobuhiro.kimura - 11:24 Thursday 16 January 2025 (32355) Print this report

{Kimura and Yasui]

 We opened gate valve called GVetmy near EYA at around 10:30 AM.
And we changed the operating threshold of the alarm contacts within CC-10 to 5.9x10^-5 Pa.
As reported K-log(32354), the alarm contacts are designed to operate when the threshold is exceeded.
We set the threshold at 0.1 lower than the previously agreed value of 6.0x10^-5 Pa.
It has been confirmed that GVetmy closes when the operating threshold is exceeded.