VAC (EX)
nobuhiro.kimura - 16:24 Wednesday 15 January 2025 (32345) Print this report
Re-enabled the interlock circuit for GVetmx

[Kimura and Yasui]

 At around 15:00, after opening GVetmx, the interlock device for GVetmx was re-enabled.
The pressure gauge for the interlock device installed in X-30 indicated 3.1 x 10^-6 Pa with GVetmx closed and 6.1 x 10^-6 Pa after GVetmx was opened.
The threshold for the interlock device to operate was set at 8.0 x 10^-6 Pa.

Comments to this report:
takahiro.yamamoto - 17:10 Wednesday 15 January 2025 (32350) Print this report
The Guardian alert was resumed with a threshold value of 7.0e-6Pa.

Readout value: ~6.0e-6Pa at 17pm
Alert threshold: 7.0e-6Pa
Interlock threshold: 8.0e-6Pa