Reports 1-1 of 1 Clear search Modify search
VAC (Valves & Pumps)
nobuhiro.kimura - 2:47 Saturday 11 January 2025 (32306) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

[Kimura and Yasui]

 BS CC-10 pressure reached 7.5 x 10^-6 (Pa) at 16:10. 

The threshold of the CC-10 contact output was then set to (7.5 + 0.5) x 10^-6 (Pa) of 8.0 x 10^-6 (Pa). 

After setting, the interlock device was enabled.

 