Reports 1-1 of 1 Clear search Modify search
MIF (General)
takaaki.yokozawa - 5:49 Thursday 26 December 2024 (32133) Print this report
Ring down violin Q meaurements
Requested by Abe-san,
With the same manner of klog31965 and klog31997, we excited the violin peaks and measured the ring down.

IX :
IM temperature 180 K
Excitation stop 2024/12/26 5:31:00(JST)
Peak frequency 174.047 -> 174.141

IY :
IM temperature 182 K
Excitation stop 2024/12/26 5:35:00(JST)
Peak frequency 176.504 -> 176.594

EY :
IM temperature 187 K
Excitation stop 2024/12/26 5:39:00(JST)
Peak frequency 187.337 -> 187.421

EX :
IM temperature 190 K
Excitation stop 2024/12/26 5:43:00(JST)
Peak frequency 184.562 -> 184.594
Comments to this report:
kentaro.komori - 0:41 Monday 06 January 2025 (32234) Print this report

I analyzed the ringdown data to estimate the violin Q-values.
The estimated Q-values are as follows:

- IX: 1.2e4
- IY: 0.86e4
- EX: 1.8e4
- EY: 2.0e4

If the sapphire fibers are completely dominated by the thermo-elastic noise, the Q-values should be 2.8e4 at 200 K.
Therefore, current Q-values are probably determined by the Brownian loss.

Images attached to this comment