BS health check was done.
Unfortunately, BF GAS was not performed.
The gain of IM V1 response seems to be several dB lower. Many peaks appears from 2 Hz to 7Hz when TM coils excited.
The others seem to be healthy.
BS health check was done.
Unfortunately, BF GAS was not performed.
The gain of IM V1 response seems to be several dB lower. Many peaks appears from 2 Hz to 7Hz when TM coils excited.
The others seem to be healthy.
Though the gain of IM V1 response is lower than the reference by 2~3dB, this situation is the same as the last health check.
After replacing the function generator for the LVDT driver, the health check was performed. The BF GAS was not performed. Q of F0 and F1 GAS were increased. The coupling from Y to V at 0.2Hz was decreased in IM. The others look healthy.
The coupling from P to V was larger than the reference in IM. This situation is the same as the last measurement. Compared with other coupling passes, the reference level is too small.