Abstract:
HPCD noise might be reduced by a factor of 5-6 by just changing the order of the engagement of dewhitening filter.
It should be checked with the test bench or real circuit.
Detail:
To investigate the noise from coil driver outputs, I checked the noise spectra of coil drivers.
First, I noticed the high power coil driver (HPCD) output noise is reduced a lot when chaging the dewhitening filter (DWF) settings from 2 to 3 stages while the noise reduction is small from 0 to 1 and 1 to 2.
So, I checked why this noise reduction happens.
This time, I used the data on the JGWdoc (JGW-S1807833-v1) for the calculation.
According to the circuit diagram (JGW-D1503503-v4), 16k-Ohm resistances at the dewhitening filter stages could limit the output noise of the HPCD.
Figure 1 shows the noise spectrum of HPCD outputs without DWF (blue) and with 3-stage DWFs (orange).
Green line in the graph shows the thermal noise of 16k-Ohm resistances at the DWFs (squrare root of square sum of 3 resistances).
Red line in the graph represents the thermal noise of 16k-Ohm resistance with single DWF because resistance at the final stage is reduced only by the lase DWF.
Since DWFs are engaged from those close to the input, final 16k-Ohm thermal noise is reduced only when all DWFs are on.
Probably that is why the noise spectrum measured in JGW-S1807833-v1 is reduced al lot only when changing DWF setting from 2 to 3.
So, if the order of engagement of DWFs is changed, HPCD output noise could be reduced.
It should be tested.