Before this work, I checked the Pcal-X beam positions on the ETMX, and adjusted a little.
What I used are PCAL_EX2 picos.
Pico movment:
- PCAL_EX2 pico 1: +50
- PCAL_EX2 pico 2: +100
- PCAL_EX2 pico 4: +50
Beam positions on the ETMX changed by about 2mm:
Before ([0,0] means the design position.): Fig 1 = TCam_ETMX_00111_2024_1120_051540_fitting_comp.png
- Pcal beam position (path1) [mm]: [3.91584, 2.40329]
- Pcal beam position (path2) [mm]: [2.75383, 3.67881]
After: Fig 2 = TCam_ETMX_00111_2024_1120_054412_fitting_comp.png
- Pcal beam position (path1) [mm]: [3.98096, -0.55771]
- Pcal beam position (path2) [mm]: [2.67615, 1.67651]
During calibration at the end station, we found the beams were slightly(2-3mm) shifted from the centre on the RxPD.
Today, we proceeded the calibration without making the adjustment. The impact on the calibration results appears to be limited, but it may need to be adjusted in the future.