[miyakawa, tanaka, ushiba, uchiyama, miyoki]
Around 12:41:21 JST, the laser stopped. The chiller for the FB laser seemed to work via webcam.
I asked Hayakawa-kun to check the water flow of the chiller for the beam dampers. The photo shows the water sending and returning of the chiller. The left is the sending and the right is the returning. When he removed the tube of the sending site, the water came from the chiller. On the other hand, they removed the tube of the returning side, then no water...
We concluded that the stuck of water happened again as klog#29363.
Miyakawa-kun and Tanaka-kun entered the PSL room and they confirmed that the interlock of the thermometer for the beam damper was triggered. They are now trying to recover from this situation.
note: some dirty material was found in the water vessel of the chiller. So we need some filter inside this water vessel to catch them. Also we need regular cleaning with some proper material.