During the repair of the water pump system in the EY area, I changed the heater/cooler/FFU operation at EYV because all FFUs and the air conditioner were operated simultaneously with heaters having a power of ~ 1kW as,
- off the air conditioner,
- off 9 of 10 FFUs
- off small heaters of ~ 600W
- on Sunrise heater at the level 2 (~ 900W).
I kept this situation for the weekend.
On the other hand, I recovered the 4 heaters (one: Delonghi, and 3 Sunrise heaters) condition at the EYC 1F area after the off of all TMP/DRY pumps because of the reoperation of the cryocoolers. However, the 3 heaters that were powered by the same voltage inverter stopped maybe because of the breaker trip of this voltage inverter (from 200V to 100V) for 3 heaters. Before, 3.3kW (1.2kW for 2 Sunrise heaters and 900W for one Sunrise heaters) was loaded in this voltage inverter. 1.5kw for the Delonghi heater was separately powered by other power route. After, I loaded 3.6kW (1.2kW for 3 heaters) while reducing the heater power from 1.5kW to 900W for the Delonghi heater. Then the breaker in the voltage inverter seemed to trip.
The temperature at EYC seemed to reduce by 1 C compared with before. I am now asking Takahashi-san about the possibility of operating the heater for the Colgate tube for the VIS adjustment.