Measurement the OLTF of the MICH and PRCL after control roll off
[Tanaka, Yokozawa]

We performed the measurement of the OLTF of the MICH and PRCL control in the state of the LOW_NOISING_PRMI (after the MICH and PRCL control roll off)
UGF of MICH was lower than before (10 Hz -> 7 Hz)
UGF of PRCL was higher than before (15 Hz -> 20 Hz)