VIS (General)
ryutaro.takahashi - 17:15 Friday 27 September 2024 (31146) Print this report
Spectra of FLDACCs

I compared the spectra of FLDACCs for Type-A in LOCKAQUISITION mode. The typical noise level is 10-210^{-2} [um/s /rtHz] at 1Hz. The H1-H3 spectra in EX are good. The H1-H3 spectra in EY are much larger than the typical level. The H2 and H3 spectra in IX are larger than the typical level at <1Hz. The H1 spectrum in IY is larger than the typical level around 1Hz.

Images attached to this report
Comments to this report:
ryutaro.takahashi - 16:05 Thursday 03 October 2024 (31186) Print this report

I compared the spectra of FLDACCs for Type-A in READY mode. The noise behavior was almost the same as the spectra in LOCKAQUISITION mode. The H1-H3 spectra in EX are good. The H1-H3 spectra in EY are much larger than the typical level (10-210^{-2}). The H2 spectrum in IX is larger than the typical level at <5Hz. The H1 spectrum in IY is larger than the typical level at <4Hz.

Images attached to this comment