Reports 1-1 of 1 Clear search Modify search
MIF (General)
Haoyu Wang - 20:31 Monday 12 August 2024 (30821) Print this report
Optics check at POP and POS

[Yuta, Haoyu]

At POS, we found the power cable for p-pol PD was missing. But we found a power cable around POP table and used it for POS p-pol PD.

The camera looking at POS s-pol does not work. There is no signal. But the camera is warm and we assume the cables are connected. We also find the mount for the camera is not stable, so we fastened the screw. The camera needs realignment. Since we don't have the image, we cannot do this.

 

At POP, the steering mirror for RST4 is BB1-E03. The label is wrong (we didn't change it).

 

Gains of the 4 PDs (POP s-pol, POP p-pol, POS s-pol and POS p-pol) are all 40dB. Whitening filters for them are all off (no gain no filter).

Comments to this report:
shoichi.oshino - 15:32 Tuesday 13 August 2024 (30826) Print this report
POS GigE camera responds to ping, but client software does not start up.
Therefore, I restarted the POS client on the camera server.
Search Help
×

Warning

×