satoru.takano - 8:08 Friday 19 July 2024 (30466) Print this report
OMC Commissioning Day3: Tapping test

Aso, Komori (remote), Takano, Washimi

Here is a brief report of the tapping test performed yesterday.

Sensitive points: the duct (Figure1) and the bottom where the pump has been located (Figure2)
Less sensitivie points: the top chamber (Figure3, Figure4), the floor smashing with a fist (Figure5)
Not sensitive points: the side hatches (Figure6), the floor close to a leg beneath OMMT1 (Figure7), the floor outside engiri (Figure8)

A short timeseries when tapping the duct is shown in Figure9. We can see an oscillation at 92.7 Hz in the feedback signsl (K1:OMC-PZT_HV1_OUT_DQ) and the error signal (K1:OMC-LSC_ERR_OUT_DQ), and the second harmonics of this can be seen in OMC TRANS.

Images attached to this report
Comments to this report:
tatsuki.washimi - 11:36 Friday 19 July 2024 (30469) Print this report

I made the tapping place and the data clear.

Images attached to this comment
tomotada.akutsu - 13:01 Friday 19 July 2024 (30470) Print this report

Hmmm... strange. In case some bolts for feet or legs of the OMC chamber would be loosen, it might be nice to ask Kimura-san to check their axial forces with torque wrenches...