Because the latest result of bruco for MICH had was provided with the channel list for DARM, the result was contaminated by lots of in-loop signals of MICH such as MICH_IN1, BS_TM_*_L, ... in all frequency bins. So I prepared a new channel list for MICH as /users/DET/tools/Bruco/share/O4/exclude_TY_MICH_240626.txt in order to provide a meaningful results. I'm now making a channel list also for PRCL as /users/DET/tools/Bruco/share/O4/exclude_TY_PRCL_240626.txt, but it's not completed yet.
For the commissioning in various IFO configuration in parallel, I also prepared a link to bruco results of latest_MICH and latest_PRCL. Though it may be better to prepare the more detailed links such as PRX, PRY, etc., such kind of implementation requires more time. Now bruco script decides link name based on the main channel name. On the other hand, it need to check guardian states at that time to decide the link name. If there is a request, it might be implemented.
-----
The `gpstime` package is expired on the main bruco server because it runs old OS. So bruco doesn't work with the input of local time or UTC. It still works with the input of GPS time. A new bruco server itself is already available from the commissioning dock (see Fig.1, "execute (new)" button is to use a new server). But it still works on DHCP, so the old server should be retired and replaced to a new server.