[Yuta, Shalika, Kenzaburo, Haoyu]
(The new layout at TMSX can be found in JGW-T1808962-v18, also shown in Fig 1.)
We replaced all BS (RBS1, RBS3 and RBS4) on the IR main beam with non-polariazing beam splitters from sigma-kk. RBS1 and RBS3 are 2-inch. RBS4 is 1-inch.
Since we replaced RBS1, we realigned the beam going to the two DCQPDs. The alignment process is done without the attenuator (the BS between RBS1 and RBS2). After the alignment, we put the attenuator back. There may be slight offsets in the yaw signals of DCQPDs due to the transmission of the BS, but should be very small.
We aligned the IR PD (RPD1), s-pol PD (RPD2), p-pol PD (RPD3) and the CCD. Reflected beams from these sensors are properly dumped.
Cannels of s-pol/p-pol PDs are added on the MEDM screen. We changed the HWP readings and checked the power changes of the two PDs. The gain of the two PDs are set as 30dB. Fig 2 shows the fitting of the power change. The HWP rotation for min Ppol(16.3) is 10.1 deg and max Spol(297) is 9.7deg. We set the rotation of HWP at 20 deg so that there is a small potion of light at p-pol PD and we are going to measure their noise performance tonorrow.
By the way, we found the IR beam height from the periscope to RBS3 is around 2mm higher than 4inch. As the mounts for the lenses RLNS1 and RLNS2 are 4-inch height, the beam is not centered for the lenses. We didn't change this.