tatsuki.washimi - 21:44 Tuesday 25 June 2024 (30109) Print this report
vibration measured for the refurbished OMC optical table

With Aogaku members,

 

We measured the vibration of the OMC optical table. See also the previous study [klog29578].

We located the 3-axial ACC on the optical table and the 1-axial ACC (for vertical) on the base plate. (Fig. 1)
Fig. 2 is the amplitude spectral densities for each channel, without injection. The dotted lines are the previous results on May 22. We can see some reductions.

After that, we performed vertical shaker injections for the base plate and evaluated the TFs (Fig. 4, 6, 8).
They look to be improved. Detail comparisons will be provided later.

Images attached to this report
Comments to this report:
tatsuki.washimi - 11:24 Wednesday 26 June 2024 (30124) Print this report

I overplot the TFs with the previous result.
The improvement is seen.
Over 400Hz may be acoustic coupling.

Images attached to this comment
tatsuki.washimi - 16:59 Wednesday 03 July 2024 (30235) Print this report

I performed shaker injection tests for the OMC baseplate in the Z-direction around 300Hz, using TEAC707 accelerometer (lower sensitivity model) and shaking larger, because significant excess was not found around 300Hz in the previous work.

Images attached to this comment
tatsuki.washimi - 17:19 Wednesday 03 July 2024 (30236) Print this report

I also tried to perform a shaker injection test horizontally (Y-direction).
It succeeded around 100 Hz, but when I injected below 50Hz, the shaker became very glitchy. 
So I gave up more investigations.

Images attached to this comment