Reports 1-1 of 1 Clear search Modify search
MIF (General)
kenta.tanaka - 19:26 Monday 17 June 2024 (29947) Print this report
RFPD phasing

[Yokozawa, Dan, Yuzurihara, K. Tanaka]

## Abstract

We performed the phasing for I/Q signals of RFPDs on POP and REFL table (POP_PDA1_RF{17, 45}, REFL_PDA1_RF56) with Frequency excitation and minimized each Q signal.

## What we did

  •  Before the phasing, we aligned PR3 to maximize the normalized GR flash to about 1, and we locked X-arm with both IR and GR beams and engaged each ADS to maximized each transmitted power. Then we recorded the oplev values of IMMT2, PR2, PRM in GOOD_OPLEV values when the transmitted powers seemed maximum.
  • We lost the lock at once and then relock X-arm with only IR beam. We connected the excitation path( DGCARM EXC -> SUMMING NODE EXC A -> SUMMING NODE SLOW OUT -> CARM CMS IN2). In that time, we decreased the CARM IN2 GAIN from 0 dB to -20 dB before connecting the path. We injected the excitation signal at 1026 Hz from K1:LSC-DGCARM_EXC ch to oscillate the laser frequency.
  • We could see a beat signal in I and Q of an POP_PDA1_RF17 at 1026 Hz. We measured the ratio between I and Q signal amplitude and obtained the rotation angle from \arctan(Q/I) . This angle indicated the phase difference between the signal and the I quadrature. So we can minimize the Q signal with putting this angle in demod. angle.
  • We rotated the demod. phase (K1:LSC-POP_PDA1_RF17_PHASE_R) to minimize the Q signal because we used POP17Q as the error signal of MICH so that the frequency oscillation would not be seen in POP17Q. Fig 1 shows the phasing result. As you can see, the Q signal are coverd by the noise level.
  • We also performed thephasing for RFPDs (POP_PDA1_RF45, REFL_PDA1_RF56) with the same procedure. Fig 2 and 3 show the result. Each Q signal got small enough to be coverd by the noise level.

## ToDo

  • We need to tune the relative gain and phase between I and Q signal for REFL_PDA3_RF45 and REFL_PDA4_RF17 with AM laser before the phasing because the I and Q signal are not always orthogonal due to the analog phase shifter error.
  • We need to perform the phasing for the 3f PDs (REFL_PDA2_RF{51, 135}) with MICH or PRCL after we succeeded MICH and PRCL lock because the beat signal between the carrier and 3f sidebands (not +1f and -2f) was too small to see the frequency oscilation. 
Images attached to this report
Comments to this report:
takaaki.yokozawa - 16:07 Tuesday 18 June 2024 (29961) Print this report
From the Today commissioning work, we suspected the phasing of the POP17 RFPD(yesterday work), so we measured again.
For this measurement, we excited the BS L (25 Hz, ~1000 cnt) from TM TEST.

As a result, we noticed there are peaks in both I and Q.
I_peak : 0.239
Q_peak : 0.116
From the calculation, we rotated from -22 deg. to -137 deg. (original phase before yesterday work...)
Then, we confirmed the I peak disappeared and only Q peaks appeared.

Conclusion : We need to take care of the IFO configuration and excitation.

Fig.1. showed the configuration of the awggui during excitation
Fig.2. showed the spectrum for (red)after the phasing (blue)no excitation (Sorry we don't have a spectrum before adjustment of the phasing)