Followings are the details of the diagonalization procedure I tried in this time.
1. Horizontal sensor balancing:
To balance the horizontal sensors, I excited the lowest yaw mode (about 0.3 Hz) and measured the ratio between H1 and H2/H3 at resonant frequency.
Then, change the sensor gain ({MN,IM}_OSEMINF_{H1,H2,H3}_GAIN) so that these ratio becomes unity (this time, I set maximum absolute gain as 1).
Figure 1 shows an example of the measurement (Top 2 graphs of right column represent IM and MN horizontal sensor signal ratio from the top).
Pink and cyan line show the ratio before gain tuning, and red and blue line represent the ratio after tuning.
2. Vertical sensor balancing:
Basic concept is same as horizontal sensor balancing but used resonance is the lowest vertical mode (about 9.4 Hz).
Figure 2 show an example of the measurement.
Same as horizontal case, top 2 figure in right colomn show the ratio of IM/MN vertical photosensors.
Pink and cyan line show the ratio before gain tuning, and red and blue line represent the ratio after tuning.
Figure 3 and 4 shows the medm screen before and after tuning the sensor gain balance, respectively.
3. Sensor decoupling from horizontal to vertical DoFs:
To reduce the horizontal to vertical sensor couplings, I excited 0.3 Hz yaw resonance and measure the ratio between yaw motion and the others at resonant frequency.
What I did is check the gain and phase of yaw to the others in TF mode measurement in diaggui and set the gain at SENSALIGN matrix (If phase is 0/180, the sign of matrix element should be negative/positive, respectively).
Lower 2 rows in fig1 shows an example of the ratio between yaw and the other DoFs after decoupling.
All seems lower than 1% that would be enough at this moment.
Since L and T motions have couplings to R and P signals, I didn't decouple these two DoFs.
4. Sensor decoupling from vertical to horizontal DoFs:
Same as horizontal to vertical, I used resonance of each DoFs (V: 9.4 Hz, P: 7.5 Hz, R: 23 Hz).
Lowest 2 rows in fig2, 5, and 6 show the ratio between V/R/P to the other DoFs, respectively.
Same as horizontal to vertical decoupling, residual coupling becomes less than 1%.
Figure 7 and 8 shows the SENSALIGN matrix before and after diagonalization.
5. Health check after decoupling:
To confirm the new SENSALIGN matrix works well, I performed healt check of MN stage (results are already shown in the original post).
Since there is no large excess and some unnecessary couplings are reduced (for example, 0.3 Hz Y2R, 7.5 Hz P2L, and so on), the decoupling seems fine.