satoru.takano - 18:57 Tuesday 14 May 2024 (29503) Print this report
Investigation of the oscillation from the common mode servo

Aoumi, Kamiizumi, Tomura, Takano

Abstract

We checked whether all the common mode servos installed in the rack are oscillating or not, and found that 6 of 9 servos are oscillating. 

Background

We noticed a strange MHz peak from GrPDHY servo (klog 29446), and first thought that there is something wrong in this specific servo. However, from the investigation of AEL specialists it was found that when the negative input of the differential driver of the common mode servo is connected to GND, the opamp used in this receiver starts oscillating at some MHz. The oscillating frequency depends on the cable connecting to the input of the servo. Unfortunately, the input signals of all the most comon mode servo already installed are such signals, and therefore it is possible for all the common mode servo to oscillate at some MHz frequency.

Works

We measured the radiated signal from the servos with handmade antennas. The target servos are:

  • LSC0 rack: GrPDHX, GrPDHY
  • ALS1 rack: PLLX, PLLY, Summing node, CARM
  • ALS0 rack: FIBX, FIBY
  • IOO0 rack: IMC

LSC0 rack

We confirmed that GrPDHX oscillates at 26.6 MHz and GrPDHY at 19.2 MHz. Figure1 and Figure2 are the signals close to GrPDHY and GrPDHX, respectively. If we disconnect the cable between each servo and the demodulator, these peaks dissapeared. Even when we turned off the power of GrPDHY and remove all the cables attached to it, we saw the 26.6 MHz oscillation if GrPHDX servo was connected to the demodulator. Therefore, the oscillation signal coupled to other circuits via the rack mounted to these circuits.

ALS1 rack

Next we checked the oscillation signal around the common mode servos installed in ALS1 rack. Figure3-Figure6 are the spectra measured around PLLX, PLLY, CARM, and Summing node. We found two dominant peaks, 16.5 MHz and 27.9 MHz. It seems that all the servos installed in ALS1 rack oscillates due to the input signal.

ALS0 rack

We moved on to ALS0 rack. Figure7 and Figure8 show the spectra around FIBX and FIBY servo. It seems that these circuits are not oscillated. This may be because the input signal to FIBX and FIBY are from DC PDs in PSL room, which GND is floated from the rack.

IOO0 rack

Finally we checked the radiated signal around IMC servo, and no oscillation peak was detected as shown in Figure9. We are not sure why this servo doesn't oscillate.

Remarks

Considering the oscillating frequency and the length of the cable connected to the input of the servo, it seems that with a 1m BNC cable the oscillation frequency is aroun 26 - 28 MHz, and with a 2m BNC cable the servo oscillates around 17 MHz.

Conclusion

We confirmed that 6 of 9 common mode servos oscillate at some MHz frequency. It is also found that the oscillation affects the other circuits on the same rack. To reduce the MHz noise in CARM servo, we have to solve the oscillation problem for all the 4 servo installed in ALS1 rack.

Future plan

For GrPDHX/Y and PLLX/Y we use a monitor signal as the input of the common mode servos, and these signals are single ended. If we replace them with differential ended signals, the oscillation may stop. One solution is to make use of the differential signals of the back plane of the demodulator/phase frequency discriminator, which is now connect to AA chassis.

 

Images attached to this report
Comments to this report:
satoru.takano - 23:01 Tuesday 14 May 2024 (29505) Print this report

I found a log related to this issue: klog 12940.

According to this log, there was a large noise at 13 MHz and to reduce it a notch filter was inserted at the input port of CARM servo as attached. However, if this noise is from the oscillation of the servo, filtering of the input doens't work well, like this log.

Images attached to this comment
takahiro.yamamoto - 10:26 Wednesday 15 May 2024 (29512) Print this report

Memo

LIGO's situation
- Input signal of the common mode servo coming from I/Q demodulator is picked up as the differential signal via TNC by using an additional breakout board.
- Circuit design of our common mode board seems to be same as the latest version of LIGO's one.
- Summing Node and CARM servo seems to be connected by a same way as our situation (negative pin is connected to GND).
- CARM FAST and IMC IN2 are connected via 2-pin lemo instead of BNC/TNC, but negative pin of 2-pin lemo for FAST OUT is connected to GND.

Useful documents
1. Schematic view of CARM/ALS: LIGO-G1500456
2. Cabling around ISC: LIGO-D1200666, LIGO-D1900511
3. I/Q demod for differential output on TNC: LIGO-D1000181
4. Common mode servo: LIGO-D0901781
5. Modifications of CMS: LIGO Wiki
6. Summing Node: LIGO-D1200151

satoru.takano - 21:52 Wednesday 15 May 2024 (29521) Print this report
satoru.takano - 20:53 Friday 17 May 2024 (29569) Print this report

I forgot to mention it.
During this work we found a air-wirering of RF components around IOO0 rack as attached. Such a connection should be avoided.

Images attached to this comment