Reports 1-1 of 1 Clear search Modify search
VIS (SR2)
ryutaro.takahashi - 15:40 Tuesday 14 May 2024 (29498) Print this report
Comment to Health check of SR2 (28369)

I checked the TFs measured in air. They look healthy. There were some differences from the reference.

  1. IP: L-T-Y coupling was changed. S/N of the measurement was improved at high frequencies.
  2. GAS: The frequency of the lowest mode was shifted to the low frequency. The Q of the third mode became low.
  3. TM: The gain from L/H3 to L was larger than the reference a little.
Images attached to this comment