ITMX went to PAY_TRIPPED state around 11:42 today due to glitches on ITMX.
To investigate the glitch source, I kept ITMX in PAY_FLOAT state but it was again tripped around 11:55 today.
The trigger of the trip is large glitch on MN V1 photosensor (fig1).
Since there was almost no effect on the other sensors, it is very likely that the sensor trouble.
Figure 2 shows the MN V1 photosensor signal from about one month ago.
T cursor shows the time when vacuum pumping was started according to klog29307.
The timing when signals went zero is DGS maintainance, so the first glitch seemed to be occured on May 2nd and somehow the glitch rate was drasticaly increased in this morning.
I also checked the signals on IOP model and confirmed the signal shape in payload model and IOP model is almost same (fig3).
So, it is probably the hardware trouble of ITMX but I haven't confirmed yet if the source is inside the vacuum chamber or not.
Considerable solusions:
I haven't confirmed the trouble happens inside the chamber or outside but I listed up the considerable solusions here in case it is inside the vacuum.
1. Open vacuum chamber and repair.
2. Change sensing matrix not to use MN V1 photosensor for pitch damping control (we could not decouple pitch and vertical but not so probematic because MNV is very stiff and dominant signal should be pitch motion).
3. Use MN OpLev for pitch control not to use MN V1 photosensor for damping controls (This situation is same as ETMY just before the earthquake, so it should work).