[Ikeda, Takahashi]
We checked the resistance and inductance of the LVDT coils for the BF damper at the feedthrough flange with the flip cable.
| Last time | This time | |
| BF H1 1-6 2-7 3-8 4-9 5-G |
11.5Ω/1037.5uH 12.2Ω/8.569mH OL OL OL |
11.9Ω/1036.5uH 12.5Ω/8.565mH OL OL OL |
| BF H2 1-6 2-7 3-8 4-9 5-G |
11.2Ω/1049.2uH 12.3Ω/8.567uH OL OL OL |
11.4Ω/1049.1uH 12.5Ω/8.562mH OL OL OL |
| BF H3 1-6 2-7 3-8 4-9 5-G |
11.6Ω/1046.3uH 12.3Ω/8.568mH OL OL OL |
11.5Ω/1046.0uH 12.5Ω/8.565mH OL OL OL |
| BF V1 1-6 2-7 3-8 4-9 5-G |
11.5Ω/1039.5uH 12.4Ω/8.561mH OL OL OL |
12.2Ω/1041.0uH 12.4Ω/8.568mH OL OL OL |
| BF V2 1-6 2-7 3-8 4-9 5-G |
11.6Ω/1021.8uH 12.5Ω/8.587mH OL OL OL |
11.4Ω/1021.4uH 12.5Ω/8.589mH OL OL OL |
| BF V3 1-6 2-7 3-8 4-9 5-G |
19.7Ω/1052.1uH 12.3Ω/8.563mH OL OL OL |
12.1Ω/1051.7uH 12.5Ω/8.557mH OL OL OL |