[Ikeda, Takahashi]
We checked the resistance and inductance of the LVDT coils for the BF damper at the feedthrough flange with the flip cable.
| Last time | This time | |
| BF H1 1-6 2-7 3-8 4-9 5-G |
11.7Ω/1039.7uH 13.3Ω/8.561mH OL OL OL |
12.0Ω/1040.6uH 12.5Ω/8.561mH OL OL OL |
| BF H2 1-6 2-7 3-8 4-9 5-G |
11.2Ω/1038.7uH 11.3Ω/8.554uH OL OL OL |
11.5Ω/1040.3uH 12.5Ω/8.552mH OL OL OL |
| BF H3 1-6 2-7 3-8 4-9 5-G |
11.3Ω/1042.7uH 12.3Ω/8.573mH OL OL OL |
11.6Ω/1041.1uH 12.5Ω/8.577mH OL OL OL |
| BF V1 1-6 2-7 3-8 4-9 5-G |
11.3Ω/1040.3uH 12.4Ω/8.553mH OL OL OL |
11.5Ω/1039.2uH 12.5Ω/8.562mH OL OL OL |
| BF V2 1-6 2-7 3-8 4-9 5-G |
11.3Ω/1016.3uH 12.8Ω/8.545mH OL OL OL |
11.5Ω/1016.3uH 12.9Ω/8.551mH OL OL OL |
| BF V3 1-6 2-7 3-8 4-9 5-G |
11.3Ω/1039.3uH 12.2Ω/8.567mH OL OL OL |
11.4Ω/1041.2uH 12.5Ω/8.253mH OL OL OL |