Reports 1-1 of 1 Clear search Modify search
VIS (SRM)
takafumi.ushiba - 10:55 Thursday 25 April 2024 (29323) Print this report
Comment to Visual inspection of SRM suspension (29317)

According to the health check in January, gain of OSEM #3 (H2) and #4 (V3) are 3 dB and 5 dB smaller than before the earthquake, respectively (fig1 and 2).

Note:

Somehow, SRM IM OSEM TFs have lower gain (2-5 dB) than before the earthquake not only in H2 and V3 but also the others (fig3: case of H3).
Further investigation is necessary.

Images attached to this comment
Search Help
×

Warning

×