VIS (EY)
takafumi.ushiba - 17:14 Thursday 04 April 2024 (29083) Print this report
Height adjustment of ETMY

[Takahashi, Ushiba]

We adjusted height of ETMY with respect to the duct center.
Height measurement method is same as ITMY measurement (klog29064).

Figure 1 and 2 show the rulers set outside of the cryostat.
Figure 3 and 4 show the rulers set inside of the cryostat.
Figure 5 and 6 show the magnet position with laser from the laser leveler, which was set at the nominal height (duct center average) of the ETMY.

Images attached to this report
Comments to this report:
ryutaro.takahashi - 22:07 Thursday 04 April 2024 (29086) Print this report

I changed the setpoint of F2 GAS from 1950 to 1623 for the height adjustment. I offloaded all GAS filters.