Date: 2024/3/12
With ShingoFujii
Abstract
- Adjusted TM position information in CAL script.
- Adjusted Pcal beam positions on ETM by using the picomotors in EXA chamber. And adjusted a little in Rx module.
- Then we could get the both Pcal beams in RxPD.
- The total optical efficiency from TxPD values to Rx value with a quick measurement was 95%, which is similler to the values in O4a (96% for path 1, and 95% for path 2)
Details
- Used an LED light to lighting ETM from EXA chamber.
- Took Tcam pictures to adjuste the ETM position on the picture. (fig1)
- Block path 2 and pass path 1 to start adjustment of the path 1.
- Used picomotors on the mirror in EXA chamber to adjust the beam position on ETM. => After this work, we could see the beam in Rx module.
- Block path 1 and pass path 2 to start adjustment of the path 2.
- Used picomotors on the mirror in EXA chamber to adjust the beam position on ETM. => After this work, we could see the beam in Rx module.
- Pass both paths, and perfomered fine adjustment of the beam positions on ETM. (fig 2) => After this work, both beams were on the RxPD port without any other alignment adjustment.
- Adjusted a little alignment for the mirrors in Rx module to make both beams on the center of the RxPD port.(fig 3)