I measured MCO TFs for health check.
Figure 1 to 7 show the TFs and they seem healthier than before.
Followings are what I noticed from the TFs:
1. Overall gain increased due to the magnet change.
2. Pitch response becomes very smooth thanks to the recovery work (klog28321).
3. P2Y coupling increased
4. All actuator signs are flipped.
Due to the sign flip of actuators, I flipped the sign of COILOUTF gains to compensate it (fig8: before, fig9: after).