I ran the automeasurement script for ITMX health check on k1ctr3 from 21:09 JST.
Data will be stored at /users/VISsvn/HealthCheck/ITMX/.
I ran the automeasurement script for ITMX health check on k1ctr3 from 21:09 JST.
Data will be stored at /users/VISsvn/HealthCheck/ITMX/.
I checked the TF results, except for TM stage. I wrote down what I noticed as follows.
At least, we need to repaire BF LVDT V2.
## IP
## GAS
## BF
## MN
## IM
I checked the cable resistance of BF LVDTs.
Results are as follows and BF V2 primary coil seems broken somewhere inside the chamber.
-Port 1-2(Primary coils): these should be around 24 ohms according to the documents
1-6 (V1): 25 ohms
2-7 (V2): overflow (more than several Mega ohms)
3-7 (V3): 24 ohms
-Port 1-4 (Secondary coils): these should be also around 24 ohms according to the documents
1-6 (V1): 24 ohms
2-7 (V2): 24 ohms
3-8 (V3): 24 ohms
Attached figures show the TF measurement results of ITMX.
All TFs seem healthy.
Note:
Similar as ETMX, MN roll TFs have a phase delay at high frequency.
This phase delay can be also seen in the reference, which was measured in vacuum at 250K.