CRY (Cryostat IX)
nobuhiro.kimura - 8:20 Friday 22 December 2023 (28051) Print this report
Comment to One od Cryocooler 50K Head showed strange temp enhencement (27720)

[Kimura and Araki (KEK)]
 The temperature of the piping in the compressor for the  duct shield cryo-cooler at BS side (Xfs) was measured.
The results are shown below.
  Row 2 is without circulator.
  Row 3 is one circulator
  Row 4 is one circulator and spot cooler added.

 We confirmed the temperature of the capillary tube and the base of the compressor is reduced by about 6°C due to the enhanced cooling.
As a result, a margin was secured against the thermal switch attached to the gas outlet piping, which is approximately 93°C..

 

20/Dec.

 9:58~10:10 21/Dec. 15:14~15:30 21/Dec.
Items without circulator one circulator Four circulators and one spot cooler added
Helium gas supply (°C) 21.9 21.3 22.7
Helium gas return (°C) 24.3 23.9 24.0
Adsober OUT (°C) 22.5 21.5 22.8
Adsober IN (°C) 21.1 21.0 22.1
Oil separator IN (°C) 20.1 19.6 19.8
Oil separator OUT (°C) 20.9 20.5 21.1
Heat exchanger He IN (IN-1) (°C) 89.1 78.3 83.9
Heat exchanger OIL IN ( IN-2) (°C) 89.2 85.5 89.0
Surface temp. of the main unit (°C) 65 65 55
Capillary tube temp. for oil return (°C) 68.8 73.3 60.7
Gas outlet piping temperature (°C) 93.4 93.5 87.6
Water inlet temp. (°C)  19.7 18.7 18.3
Water outlet temp. (°C) 25,1 24.5 24.3
Psupply (MPa) 2.28~2.32 2.28~2.32 2.30~2.33
Preturn (MPa) 0.648~0.689 0.662~0.738 0.655~0.710
Pdeff (MPa) 1.63~1.63 1.618~1.582 1.645~ 1.620