[Ushiba, Capote, Nakano]
We tried to lock the PRFPMI, but we found that the ETMX has totally different actuator response from the one measured in August, due to warming up from 30K to 220K. That would cause the 0.7Hz oscillation and can be predicted to have many other problems. So, we decided not to keep trying it, but to start the re-diagonalization script before further trial.
Activity log
1010 Found that the BIO setting of ETMY was wrong. The TM input is switched to the test input. This was for Q measurement, and we forgot to bring it back. We restore it and check all other suspensions as well.
1019 Measured the ETMY actuator response with the ALS DARM, locking only ALS CARM. The actuator response was the same as in August.
1030 Earthquake
1200 We found the BIO setting is not causing the 0.7Hz oscillation. We still have the same issue as yesterday, we can't hold at ALS_DARM due to the 0.7Hz oscillation. We can't lock
1508 Try 0.7Hz lead filter implemented into ETMY LOCK filter. No good.
1514 Try 0.7Hz lead filter implemented into ETMX LOCK and ETMY LOCK filter. No good.
1518 Try 0.7Hz lead filter implemented into ETMX LOCK filter. No good.
1605 Measured the ETMX actuator response of the TM stage with the ALS DARM, locking only ALS CARM. We found the actuator response has changed from August. Seems to have a big change at 0.6Hz peak and 1-2Hz structure.
1605 Measured the ETMX actuator response of the IM and MN stage with the ALS DARM, locking only ALS CARM. Big difference, and we feel we need to re-diagonalize the actuators.
1630 Start diagonalization script.