PEM (Center)
takaaki.yokozawa - 15:27 Tuesday 18 July 2023 (26011) Print this report
Tapping test OMC chamber
[YokozaWashimi]

We performed the tapping test for OMC chamber including the geophone.

Red : Geophone
K1:PEM-ACC_OMC_VACTABLE_OMC_Y_OUT (dush : reference)

Blue : Accelerometer on the OMC chamber (Fig.1.)
K1:PEM-PORTABLE_OMC_BOOTH_OMMT_BNC1_OUT (dush : reference)

As you can see, even there are some differences, noise level became larger when tapping the OMC chamber.
And the noise level larger when we tapped the side of chamber than top of chamber.
Images attached to this report
Non-image files attached to this report
Comments to this report:
tatsuki.washimi - 16:14 Tuesday 18 July 2023 (26012) Print this report

We performed the shaker injection (swept, band-limited white) for the berrows leg (-Y side).
Fig.1 is the spectrogram (ASD) and Fig.2 is relative to the reference.

The response looks linear below 160 Hz.
The high-frequency response is up-converted. (due to the table? or the geophone?)

Images attached to this comment
tatsuki.washimi - 17:58 Tuesday 18 July 2023 (26014) Print this report

I checked the spectrum, coherence, and transfer function for the 60-200Hz shaker injection data.
Note that the direction is different.

 

Images attached to this comment