MIR (EY)
Masayuki Yamamoto - 7:36 Monday 10 July 2023 (25916) Print this report
Detailed results of analysis of Q-value measurement in elastic vibration mode of ETMY

M.Yamamoto

On June 28, under FPMI lock.
We tried to excite the elastic vibration mode of ETMY and measure its peak frequency and decay time.
As a result, we confirmed the resonance which was estimated to be derived from five elastic vibration modes.

We also confirmed that each resonance decayed within about 0.1 to a few seconds after the excitation signal was turned off.
Q-values determined from the decay of the resonances are shown here.

https://klog.icrr.u-tokyo.ac.jp/osl/?r=25771 Brief report by Mr.Komori.

Q-value were recorded ten times for each resonance, and the average and standard error of Q-value were calculated

(however, 51.0367 kHz was recorded only nine times).
 I will upload the detailed data of to JGW-doc later.
 
Attention:Due to an analytical error, the Q values of ITMX and ITMY measured in the past may be slightly different(at most twice as much).
So I will reload the corrected data on k-log later.

Q value at each resonant frequency
Resonant frequency[kHz] Q-value(average) Q-value(SE)
23.7124 2.8E+04 2.2E+03
34.4963 2.4E+05 1.0E+04
50.7746 9.9E+04 5.0E+02
51.0367 3.0E+4 1.0E+03
51.1698 2.0E+04 6.0E+02

 

Images attached to this report