M.Yamamoto
On June 28, under FPMI lock.
We tried to excite the elastic vibration mode of ETMY and measure its peak frequency and decay time.
As a result, we confirmed the resonance which was estimated to be derived from five elastic vibration modes.
We also confirmed that each resonance decayed within about 0.1 to a few seconds after the excitation signal was turned off.
Q-values determined from the decay of the resonances are shown here.
https://klog.icrr.u-tokyo.ac.jp/osl/?r=25771 Brief report by Mr.Komori.
Q-value were recorded ten times for each resonance, and the average and standard error of Q-value were calculated
(however, 51.0367 kHz was recorded only nine times).
I will upload the detailed data of to JGW-doc later.
Attention:Due to an analytical error, the Q values of ITMX and ITMY measured in the past may be slightly different(at most twice as much).
So I will reload the corrected data on k-log later.
Resonant frequency[kHz] | Q-value(average) | Q-value(SE) |
23.7124 | 2.8E+04 | 2.2E+03 |
34.4963 | 2.4E+05 | 1.0E+04 |
50.7746 | 9.9E+04 | 5.0E+02 |
51.0367 | 3.0E+4 | 1.0E+03 |
51.1698 | 2.0E+04 | 6.0E+02 |