CAL (General)takahiro.yamamoto - 22:05 Wednesday 21 June 2023 (25666)
Print this reportMeasurements for the post O4a calibrationWe did the full measurements (JGW-L2314903 as the post O4 calibration.
Quick analysis is consistent with measurements of pre-O4 calibration klog#25024. ETMX_TM / Pcal with DARM: (-3.4492 +/- 0.0097)e-12 ETMX_IM / ETMX_TM with DARM: 40.58 +/- 0.31 ETMX_IM: (-1.400 +/- 0.011)e-12 ITMX_TM / ETMX_TM with DARM: 0.8261 +/- 0.0018 ITMX_TM: (-2.849 +/- 0.010)e-12 BS_TM / ITMX_TM with mich: 21.288 +/- 0.066 BS_TM: (-6.065 +/- 0.029)e-11 PRM_TM / BS_TM with PRCL: 27.69 +/- 0.36 PRM_TM: (-8.39 +/- 0.12)e-10
Largest changes is ETMX_TM as ~2.2%. Weekly measurements also have a 2-3% fluctuation. So estimated values seem to be reasonable.