Operators name: Yuzurihara, Ikeda
Shift Time: 08-16
Summary:
- Check Items
=> No Probrem - 10:29-11:28 PR3 Alignment(Oshino, Yuzurihara -> Ikeda)
- 12:26 lock loss.due to loading guardian during lock
- 12:44 observation
Operators name: Yuzurihara, Ikeda
Shift Time: 08-16
Summary:
> 12:26 lock loss.due to loading guardian during lock
I thought that just (re-)loading Guadian itself might not directly affect the interferometer control... no?
>I thought that just (re-)loading Guadian itself might not directly affect the interferometer control... no?
It depends on which guardian is reloaded and how reload is done.
This time initial alignment guardian was reloaded by once stop it.
So, the main function of its DOWN state was run, which requests DOWN state to XARM, YARM, VERTEX, and ASC_LOCK guardians.