Reports 1-1 of 1 Clear search Modify search
VAC (General)
takaaki.yokozawa - 8:58 Monday 22 May 2023 (25323) Print this report
Many vacuum monitor value zero
I found many vacuum monitor value is zero
It might start from 2023/05/20 16:00:00 (Sat)

http://172.16.33.39/vacuum/vac3.html
Images attached to this report
Comments to this report:
takaaki.yokozawa - 10:44 Monday 22 May 2023 (25325) Print this report
YamaT-san checked the network, we could not find the problem.
I also checked the error log, if the connection of ssh and/or serial transportation between raspberry pi and cc-10 was failed, the log remained, but no log.
So we restarted the script and keep monitoring them.
takaaki.yokozawa - 8:09 Tuesday 23 May 2023 (25335) Print this report
After restarted the software, the situation became better, but sometimes it occurred.
This is a software problem, not the hardware problem in this case.
Anyway, if you noticed the vacuum level sometimes became zero frequently, please contact to Yokozawa soon.
Images attached to this comment