Reports 1-1 of 1 Clear search Modify search
PEM (Center)
tatsuki.washimi - 16:39 Friday 19 May 2023 (25311) Print this report
Comparison of the acoustic injection results at the REFL area

I compared the results of the swept sine acoustic injection tests at the REFL area, performed on May 2 and 16.  (Note that I have not checked the calibration consistency between these days.)
The first figure is the 0-1000 Hz range and the others are their zooming up.

  • 150 counts (blue) vs. 200 counts (orange) of the DAC output, 50-900 Hz scan with 5 Hz steps
    • The blue and the orange are almost similar for all frequencies.
  • Around 120 Hz, 220 Hz, and 340 Hz: The current sensitivity looks to be occupied by this acoustic noise.

 

Images attached to this report