Reports 1-1 of 1 Clear search Modify search
MIF (General)
takahiro.yamamoto - 12:40 Sunday 14 May 2023 (25228) Print this report
Comment to CPU load reduction of k1lsc (25212)
Glitch situation seems to be changed after this model modification.
But it hasn't been solved yet...

Only 2 events were detected on Saturday (attachment 1).
But for today, 64 events were detected until noon (attachment 2).
Today's glitchy time seems to be finished.

In previous investigation e.g. klog#24949, we constantly detected ~30 events per day.
So the situation seems to be changed. But total number of glitches may not be changed as ~60 events per 2-days.

Because the duration of glitchy time didn't changed from ~1hour,
the total dead time was reduced from 1hour per day to 1hour per 2-days.
Non-image files attached to this comment
Search Help
×

Warning

×