tatsuki.washimi - 17:55 Tuesday 25 April 2023 (24958) Print this report
Tapping tests on EX/EY/IY

[X-end: Yokozawa, Yamazaki]
[Y-end: Washimi, Murakoshi]
[Control room: Ushiba, Yuzurihara, Suzuki, Ishikawa]
[Remote: Akutsu, Tomaru, and many people]

Today we performed the tapping test mainly X/Y-end and around IY.

We found the following excess (see also the plots).

  • TMSY (Leg, Table): 70-150 Hz
  • TMSX (Leg): ~140 Hz, ~280 Hz
  • TMSX (Cover): ~70 Hz, ~110Hz
  • EXT (Bellows leg): ~140 Hz
  • EYT (Bellows leg): ~80 Hz
  • Beam duct between IYC and BS: ~10 Hz, 30-40 Hz, ~70Hz

 

There was no  excess at

  • EXT/EYT chambers body
  • around EXC/EYC/IYC chambers (body, cryocoolers, ground, large flanges, booth frame, booth 2F, bellows, beam duct)
  • around EXA/EYA/IYA chambers (body, bellows, legs, ground, side flanges, beam duct)
Images attached to this report
Comments to this report:
takahiro.yamamoto - 9:16 Wednesday 26 April 2023 (24960) Print this report
Attachment shows a list of event time of IPC glitches during 4/25 9:00 JST - 4/26 9:00 JST.
It may help offline analyses.
Non-image files attached to this comment
takaaki.yokozawa - 9:33 Wednesday 26 April 2023 (24961) Print this report
I checked the PD and QPD signals during the tapping test at TMSY (Sorry, no ACC on the TMSY optical table)
No significant excess detected.
Images attached to this comment
takahiro.sawada - 19:04 Thursday 27 April 2023 (24988) Print this report

In most cases, IPC glitches occurred one sample before GPS seconds were divisible by 5.