Reports 1-1 of 1 Clear search Modify search
MIF (General)
Shunsei Yamamura - 18:56 Saturday 22 April 2023 (24930) Print this report
Comment to Morning silent run 230422 (24924)

I checked the Gaussianty, and I found that there are many Non-Gaussian times in the latter silent run.
(https://gwdet.icrr.u-tokyo.ac.jp/~controls/Gauch/day/20230422/index.html )
 
Because the Non-Gaussians appear in all frequencies, I thought these are similar phenomena to klog24795.

In fact, I found the whitend timeseries data has many peaks (Fig.1) and there are gaps in timeseries data (Fig.2 ~ 4 ) as same as the above report. (I investigated only three gliches from the first. I will examine more later.)

However, it seems a bit different because of the following reasons.

1. These phenomena occurred many times in a short period.

2. The gap occurred twice in each of them(after the first gap, it seems the data almost recovers with the second gap).

I don't know if these are the same as  klog24795, so I hope someone helps me.

Images attached to this comment
Search Help
×