DetChar (General)shoichi.oshino - 16:18 Wednesday 29 March 2023 (24595)
Print this reportRestarting FscanI restarted the Fscan process. Now results of Fscan can be seen from SummaryPage.
Detail: Fscan is working on k1det0 once a day with cron. Currently, Fscan uses SIENCE_MODE segments for its process. But the science mode has been not turned on during commissioning period, I switched segments list from SIENCE_MODE to LSC_LOCKED. I also modified the channel name of the strain to K1:CAL-CS_PROC_DARM_STRAIN_DBL_DQ. Fscan also generates plots of CAL-CS_PROC_DARM_DISPLACEMENT and CAL-CS_PROC_IMC_FREQUENCY channels. Magnetometer channels are used in LIGO. If these channels are needed in KAGRA in the future, we will add them to Fscan. I also checked the newer version of FscanDriver on LIGO GitLab but it doesn't work well. Maybe the options of lalpulser_MakeSDFDAG are different from before so we need to modify that.