PEM (Center)takaaki.yokozawa - 9:33 Wednesday 29 March 2023 (24584)
Print this reportTapping test at the PSL room[YokozaWashimi with Akutsu(remote)]
We performed the tapping test at the PSL room.
1. Tapping the optical table at various places We touched the optical table with various place. One large signal was detected when we touched around new high power laser(Fig.1. in URL) In other part, some small signal detected, but not related in existing peak
2. Tapping the leg of the optical table There are some signal, but not related in existing peak
3. Touching the cable for accelerometer on the output periscope With the advise of Akutsu-san, we touched the cable for the accelerometer on the output periscope, but no signal detected.
4. Rack for laser, electronics No signal was detected
5. Beam duct for output When we touched the beam duct toward the output to PSL room, very large signal detected (Fig.2. in URL) When we injected the white vibration using vibrator, the excess signal became smaller, that imply the low frequency signal up-convert to high frequency(Fig.3. in URL)
takaaki.yokozawa - 12:59 Wednesday 29 March 2023 (24593)
Print this reportFirst of all, I checked the IMMT1 Trans QPD2 channels. Obviously, the signal of both pit and yaw became larger and only pitch have coherence