PEM (Center)takaaki.yokozawa - 15:34 Friday 24 February 2023 (24140)
Print this reportAdditional test of vibration check near the OMC areaI continuously measured the vibration around the OMC area. The all data summary will appear soon.
1. Fun near the SR cooler on/off(Fig.1.) There is fun to circulate the air flow near the SR cooler. Currently it was turned off. I tried to turn on and off and compare the spectrum
2. SR cooler on/off (Fig.2.) I put the accelerometer on the frame under the SR cooler and turned on and off the SR cooler.
3. Vibration injection test (Fig.3.) I injected the vibration by mu leg. Check the transfer function and coherent
4. FFU(OMC/SRM/SR3) on/off (Fig.4.) Four FFUs on in each area (OMC, SRM, SR3) We turned off the all FFUs and compared the spectrum All frequency above 100 Hz was reduced of factor two and disappeared several peaks. Especially, 120 Hz reduction is so much! (But increased in 180 Hz??) Anyway, the vibration level below 200 Hz is very close to O3GK one.
Images attached to this report
Comments to this report:
tatsuki.washimi - 10:13 Monday 27 February 2023 (24168)
Print this report
I estimated the OMC ground vibration noise on DARM based on the FFU off data on 2023-02-24.
Images attached to this comment
takaaki.yokozawa - 8:03 Tuesday 28 February 2023 (24188)
Print this reportI showed the spectrum of accelerometers for this measurement. Red spectrum is always OMC leg. Coherence is between OMC leg and acc in each place.
Fig.1. mini rack near the OMC Fig.2. decicator near the OMC Fig.3. silver rack near the OMC Fig.4. OMC0 digital rack Fig.5. FFU controller for OMC area