Abstract:
We tried to activate twelve ion pumps installed on X-arm, Y-arm, IXV and IYV from 17/Jan./2023 to 20/Jan./2023.
The operating status on 20/Jan./2023 of each pump we started up is recorded on the K-log.
Nine of the twelve ion pumps were successfully activated. The remaining three pumps could not be activated due to failure of high-voltage power supply.
The causes of the failure were presumed to be a high-voltage leakage due to humidity.
See the note below for the locations of the ion pumps were not activated.
Workers:
N. Kimura
H. Yasui
M. Takahashi
X-arm:
X-27 (Position: 2700 m):
Status: Not activated, evacuated by TMP and Dry-pump
12:58, 20/Jan./2023
The pressure displayed on the pressure gage: 2.3 x 10^-6 Pa.
The ion pump installed at X-27 could not be sactivated due to a high voltage power failure.
Therefore, vacuum evacuation at the X-27 point is continued by TMP and Dry-pump. See attached photos,
X-21 (Position: 2100 m):
Status: Activated, evacuated by Ion Pump
13:05, 20/Jan./2023
The pressure displayed on the pressure gage: 1.8 x 10^-6 Pa.
Pressure displayed on the Ion pump power supply: 2.0 x 10^-6 Pa.
The anode voltage and current displayed on the Ion pump power supply : 6.1 kV and 299 x 10^-6 A.
The power consumption of the Ion pump power supply: 16 VA. (Single phase AC 200 V, current 0.08 A
X-15 (Position: 1500 m):
Status: Activated, evacuated by Ion Pump
13:12, 20/Jan./2023
The pressure displayed on the pressure gage: 1.0 x 10^-6 Pa.
Pressure displayed on the Ion pump power supply: 3.8 x 10^-6 Pa.
The anode voltage and current displayed on the Ion pump power supply: 6.1 kV and 560 x 10^-6 A.
The power consumption of the Ion pump power supply: 16 VA. (Single phase AC 200 V, current 0.08 A)
Noticed:
Dew condensation inside the box that houses the high-voltage power supply.
X-10 (Position: 1000 m):
Status: Activated, evacuated by Ion Pump
13:20, 20/Jan./2023
The pressure displayed on the pressure gage: 1.2 x 10^-6 Pa.
Pressure displayed on the Ion pump power supply: 5.5 x 10^-6 Pa.
The anode voltage and current displayed on the Ion pump power supply: 6.1 kV and 820 x 10^-6 A.
The power consumption of the Ion pump power supply: 22 VA. (Single phase AC 200 V, current 0.11 A)
X-2 (Position: 200 m):
Status: Activated, evacuated by Ion Pump
13:27, 20/Jan./2023
The pressure displayed on the pressure gage: 2.4 x 10^-6 Pa.
Pressure displayed on the Ion pump power supply: 1.0 x 10^-6 Pa.
The anode voltage and current displayed on the Ion pump power supply: 6.2 kV and 157 x 10^-6 A.
The power consumption of the Ion pump power supply: 18 VA. (Single phase AC 200 V, current 0.19 A)
Noticed:
Dew condensation inside the box that houses the high-voltage power supply.
Y-arm:
Y-27 (Position :2700 m):
Status: Activated, evacuated by Ion Pump
The pressure displayed on the pressure gage: 3.0 x 10^-6 Pa.
Pressure displayed on the Ion pump power supply: 1.8 x 10^-6 Pa.
The anode voltage and current displayed on the Ion pump power supply: 6.1 kV and 274 x 10^-6 A.
The power consumption of the Ion pump power supply: 22 VA. (Single phase AC 200 V, current 0.11 A)
Y-21 (Position: 2100 m):
Status: Activated, evacuated by Ion Pump
10:30, 20/Jan./2023
The pressure displayed on the pressure gage: 1.4 x 10^-6 Pa.
Pressure displayed on the Ion pump power supply: 1.9 x 10^-5 Pa.
The anode current displayed on the Ion pump power supply : 600 x 10^-6 A.
The power consumption of the Ion pump power supply: 26 VA. (Single phase AC 200 V, current 0.13 A)
Y-15 (Position: 1500 m):
10:36, 20/Jan./2023
Status: Not activated, evacuated by TMP and Dry-pump
The pressure displayed on the pressure gage: 7.7 x 10^-7 Pa.
The ion pump installed at Y-15 could not be activated due to a high voltage power failure.
Therefore, vacuum evacuation at the Y-15 point is continued by TMP and Dry-pump.
Y-10 (Position: 1000 m):
Status: Not aActivated, evacuated by TMP and Dry-pump
10:56, 20/Jan./2023
The pressure displayed on the pressure gage: 7.7 x 10^-7 Pa.
The ion pump installed at Y-10 could not be activated due to a high voltage power failure.
Therefore, vacuum evacuation at the Y-10 point is continued by TMP and Dry-pump.
Y-2 (Position: 200 m):
Status: Activated, evacuated by Ion Pump
11:07, 20/Jan./2023
The pressure displayed on the pressure gage: 2.0 x 10^-6 Pa.
Pressure displayed on the Ion pump power supply: 1.5 x 10^-6 Pa.
The anode voltage and current displayed on the Ion pump power supply: 6.1 kV and 215 x 10^-6 A.
The power consumption of the Ion pump power supply: 14 VA. (Single phase AC 200 V, current 0.07 A)
Center area:
IXV (Position: Center 2F):
Status: Activated, evacuated by Ion Pump
14:13, 20/Jan./2023
The pressure displayed on the pressure gage: 5.9 x 10^-6 Pa
The pressure displayed on the Ion pump power supply: 4.9 x 10^-6 Pa
The anode voltage and current displayed on the Ion pump power supply: 6.3 kV and 490 x 10^-6 A
IYV (Position: Center 2F):
Status: Activated, evacuated by Ion Pump
14:13, 20/Jan./2023
The pressure displayed on the pressure gage :1.0 x 10^-5 Pa
The pressure displayed on the Ion pump power supply: 6.1 x 10^-6 Pa
The anode voltage and current displayed on the Ion pump power supply: 6.2 kV and 921 x 10^-6 A