After today's Kimura-san's vacuum leak repair in the OMC and OMMT tank, TMP was successfully turned on (This report should be done by Kimura-san).
However, the vacuum level stopped at 0.12 Pa suddenly and kept constant.
I re-entered the KAGRA center area and I confirmed that the TMP/Scroll pumps were working. So actually other leaks should exist.
I tried the "ethanol method", and found a leak at the GV-side d400 flange on the OMMT tank because the vacuum level enhanced from 0.12 Pa to 0.6 Pa when I put ethanol. Also I found several suspicious blank flanges and a connector flange on the OMMT and OMC tank.
I hope they will be repaired by Kimura-san.